PHY 315 · Electronic Instrumentation I

Static performance characteristics and calibration

Accuracy, precision, sensitivity, linearity, hysteresis, resolution, threshold, drift, range, and fitting a calibration line.

Our notes

2 min read

Static characteristics

How an instrument behaves when the input is steady or changes slowly.

  • Range: the lowest to highest values it can measure (0-100 °C). Span: highest minus lowest (100 °C).
  • Accuracy: closeness of the reading to the true value; often given as ± % of full scale or of reading.
  • Precision (repeatability): closeness of repeated readings to each other. An instrument can be precise but inaccurate (a consistent offset).
  • Sensitivity: change in output per change in input (the slope of the calibration curve), for example 41 μV/°C for a type K thermocouple near room temperature.
  • Linearity: how closely the calibration curve follows a straight line; non-linearity is the largest deviation from the best straight line, as a % of full scale.
  • Hysteresis: different outputs for the same input depending on whether the input is rising or falling (friction, magnetic and elastic effects).
  • Resolution: the smallest input change that gives a detectable output change.
  • Threshold: the smallest input that produces any output from zero.
  • Dead zone (dead band): a range of input with no output change.
  • Drift: slow change of output with time while the input is constant: zero drift (offset) and sensitivity (span) drift, often caused by temperature or ageing.
  • Static error = measured value − true value; correction = −error.

Calibration

Calibration compares an instrument with a more accurate standard over its range and records the relation (and the uncertainty). It should be repeated at intervals and is traceable to national standards (units topic).

Fitting a straight line by least squares

Calibration data (input in °C, output in mV):

x010203040
y0.11.11.93.23.9

Slope m = [nΣxy − ΣxΣy]/[nΣx² − (Σx)²]; intercept c = (Σy − mΣx)/n.

  • Σx = 100, Σy = 10.2, Σxy = 0 + 11 + 38 + 96 + 156 = 301, Σx² = 3,000, n = 5.
  • m = (5 × 301 − 100 × 10.2)/(5 × 3,000 − 10,000) = (1,505 − 1,020)/5,000 = 0.097 mV/°C (the sensitivity).
  • c = (10.2 − 0.097 × 100)/5 = 0.5/5 = 0.10 mV (the zero offset).
  • Residuals (y − line): 0.00, +0.03, −0.14, +0.19, −0.08. Largest 0.19 mV. Full-scale output ≈ 0.1 + 0.097 × 40 = 3.98 mV, so non-linearity ≈ 0.19/3.98 ≈ 4.8% of full scale.

Check: the residuals add to zero (0 + 0.03 − 0.14 + 0.19 − 0.08 = 0), as they must for a least-squares line ✓.

Loading effect

Connecting an instrument takes energy from the system and changes the quantity measured (a voltmeter drawing current; a thermometer cooling a small sample). Good instruments have high input impedance for voltage and low impedance for current measurement.

Exams usually ask

  • Define the static characteristics of an instrument.
  • Distinguish accuracy from precision, and threshold from resolution.
  • Fit a calibration line by least squares and find sensitivity and non-linearity.

Common mistake

Treating a high-resolution digital display as proof of accuracy. Six digits can show a precise but wrong number if the instrument is out of calibration.

Written by the UniHive team in our own words. Spotted a mistake? Tell us on WhatsApp.