Functional elements

Any instrument can be described by a chain of functional elements:

  1. Primary sensing element (sensor): in contact with the measurand; responds to it (a thermocouple junction, a Bourdon tube, a strain gauge, a photodiode).
  2. Variable conversion element (transducer): changes the signal into a more useful form, often electrical (displacement to voltage).
  3. Variable manipulation element: changes the size of the signal without changing its nature (an amplifier, an attenuator).
  4. Data transmission element: carries the signal to a distant point (cables, 4-20 mA loops, radio, fibre).
  5. Data presentation element: displays or records the value (a pointer and scale, a digital display, a chart recorder, a computer screen).

Signal conditioning (filtering, linearising, converting to digital) may come anywhere in the chain.

Example: a pressure gauge (Bourdon type)

Pressure → Bourdon tube (sensing: pressure to tip displacement) → linkage and gears (manipulation: amplify the motion) → pointer and dial (presentation). No transmission element (a local gauge).

Example: an electronic thermometer

Temperature → thermistor (sensing) → bridge circuit (conversion: resistance to voltage) → amplifier (manipulation) → ADC and display (presentation).

Active and passive transducers

  • Passive (self-generating): produce their own output energy from the measurand (thermocouple, piezoelectric crystal, photovoltaic cell).
  • Active (modulating): need an external power supply, which the measurand modulates (strain gauge, thermistor, LVDT).

Input-output configuration

An instrument's output depends on three kinds of input:

  1. Desired input (i_D): the quantity we want to measure.
  2. Interfering inputs (i_I): quantities to which the instrument responds in addition (they add to the output). Example: a strain gauge also responds to temperature; mains hum picked up by leads.
  3. Modifying inputs (i_M): quantities that change the relation between desired input and output (change the sensitivity or zero). Example: supply voltage changes the output of a bridge; temperature changes a spring's stiffness in a pressure gauge.

Methods of correcting for unwanted inputs

  1. Inherent insensitivity: choose materials and designs that respond little to the unwanted input (low-temperature-coefficient alloys such as constantan in strain gauges).
  2. High-gain feedback: make the output depend on stable feedback elements (see the feedback topic).
  3. Signal filtering: remove unwanted frequency components (filters for mains hum).
  4. Opposing inputs (compensation): add a second element that responds to the unwanted input in the opposite sense: a dummy strain gauge in the next arm of the bridge cancels temperature effects; cold-junction compensation in thermocouples.
  5. Calculated output corrections: measure the interfering input and correct mathematically (common in digital instruments).
  6. Shielding and isolation: screened cables, earthing, magnetic shielding.

Exams usually ask

  • Describe the functional elements of a generalised measuring system with an example.
  • Distinguish desired, interfering and modifying inputs, with examples.
  • Describe methods of correcting for interfering and modifying inputs.

Common mistake

Calling every unwanted input "interference". Decide whether it adds to the output (interfering) or changes the sensitivity (modifying); the corrections differ.