PHY 315 · Electronic Instrumentation I

Dynamic characteristics of instruments

Zero-, first- and second-order instruments, step and ramp responses, time constant, damping, overshoot and settling time.

Our notes

2 min read

Why dynamics matter

When the input changes quickly, an instrument's reading lags behind or overshoots. Dynamic characteristics describe this: speed of response, fidelity (faithful reproduction) and dynamic error.

Order of an instrument

The relation between output y and input x is modelled by a linear differential equation; its order classifies the instrument.

Zero order: y = Kx

Output follows input instantly (an ideal potentiometer displacement sensor). K is the static sensitivity.

First order: τ dy/dt + y = Kx

  • τ, the time constant: sets the speed of response. Examples: a thermometer or thermocouple in a fluid, an RC circuit.
  • Step response: y = Kx₀(1 − e^(−t/τ)). After τ, 63.2% of the final value; after 2τ, 86.5%; 3τ, 95%; 5τ, over 99%.

#### Example
A thermometer with τ = 10 s is moved from 25 °C into 100 °C water. After 20 s:
T = 25 + 75(1 − e^(−2)) = 25 + 75 × 0.8647 ≈ 89.8 °C.
Time to read within 1% of the change (99%): t = τ ln(100) ≈ 4.6τ = 46 s.

  • Ramp response: for an input rising at a steady rate R, the output lags by a time τ, so the steady dynamic error = Rτ. A temperature rising at 2 °C/s read by this thermometer reads 20 °C low.
  • Frequency response: amplitude ratio 1/√(1 + (ωτ)²), phase lag tan⁻¹(ωτ).

Second order: (1/ωₙ²) d²y/dt² + (2ζ/ωₙ) dy/dt + y = Kx

  • ωₙ, the undamped natural frequency, and ζ, the damping ratio. Examples: moving-coil meters, spring balances, accelerometers, pressure sensors with fluid lines.
  • Step response depends on ζ:
  • ζ < 1 underdamped: oscillates about the final value with decreasing amplitude.
  • ζ = 1 critically damped: fastest response without overshoot.
  • ζ > 1 overdamped: slow, no overshoot.

  Many instruments are designed for ζ ≈ 0.6-0.7: a small overshoot but quick settling.

  • Formulas (underdamped):
  • Damped frequency ω_d = ωₙ√(1 − ζ²).
  • Peak time t_p = π/ω_d.
  • Percentage overshoot = 100 e^(−πζ/√(1 − ζ²)).
  • Settling time (2% band) ≈ 4/(ζωₙ).

#### Example
ωₙ = 100 rad/s, ζ = 0.5:

  • ω_d = 100 × 0.866 = 86.6 rad/s; t_p = π/86.6 ≈ 0.036 s.
  • Overshoot = e^(−π × 0.5/0.866) = e^(−1.814) ≈ 16.3%.
  • Settling time ≈ 4/(0.5 × 100) = 0.08 s.

Check: the envelope e^(−ζωₙt) at t = 0.08 s is e^(−4) ≈ 0.018, under 2% ✓.

Other dynamic terms

  • Rise time: time from 10% to 90% of the final value.
  • Response time / settling time.
  • Bandwidth: the frequency range over which the instrument responds well (amplitude ratio above 0.707, the −3 dB point).
  • Dead time (transport lag): a pure delay (fluid flowing along a sampling line).

Exams usually ask

  • Derive and sketch the step response of a first-order instrument; define the time constant.
  • Calculate readings and dynamic errors for first-order instruments.
  • Describe the effect of damping on a second-order instrument; calculate overshoot and settling time.

Common mistake

Saying a first-order instrument reaches the final value after one time constant. After τ it has covered only 63.2% of the change.

Written by the UniHive team in our own words. Spotted a mistake? Tell us on WhatsApp.