PHY 351.2 · Electronics I

Electronic measurements and instruments

Measurement errors, meters, the multimeter and its loading, the oscilloscope, signal generators and good practice.

Our notes

2 min read

Measurement terms

  • Accuracy: how close a reading is to the true value.
  • Precision: how close repeated readings are to each other.
  • Resolution: the smallest change the instrument can show.
  • Sensitivity: output change per unit input (for analogue meters, ohms per volt).
  • Errors: gross (human: wrong range, misreading), systematic (zero error, calibration, loading) and random (noise; reduced by averaging).

Percentage error = (measured − true) ÷ true × 100.

Multimeters

  • Analogue (moving-coil) meter: a coil turning in a magnetic field; shunts extend current ranges, series multipliers extend voltage ranges. Sensitivity typically 20 kΩ/V.
  • Digital multimeter (DMM): converts the input to a number with an ADC; input resistance usually 10 MΩ on voltage ranges; shows d.c., a.c. (r.m.s.: "true r.m.s." meters are accurate for non-sine waves), resistance, diode test, continuity, sometimes capacitance and frequency.

Loading effect

A voltmeter draws current, which changes the circuit it measures. Its resistance appears in parallel with the part being measured.

Worked example

Two 100 kΩ resistors divide 10 V, so the true voltage across the lower one is 5 V.

  • Analogue meter on the 10 V range at 20 kΩ/V: R_m = 200 kΩ. 100k ∥ 200k = 66.7 kΩ; reading = 10 × 66.7 ÷ 166.7 = 4.00 V (20% low).
  • DMM with 10 MΩ: 100k ∥ 10M = 99.0 kΩ; reading = 10 × 99.0 ÷ 199.0 = 4.98 V (0.5% low).

Extending a meter

A movement of 1 mA full scale and 100 Ω resistance:

  • To read 10 V: series multiplier R = 10 ÷ 0.001 − 100 = 9,900 Ω.
  • To read 1 A: shunt R = (0.001 × 100) ÷ (1 − 0.001) = 0.1001 Ω.

The cathode ray oscilloscope (CRO) and digital oscilloscope

Shows voltage (vertical) against time (horizontal). Controls: volts/div, time/div, trigger (to hold the trace still), coupling (a.c., d.c., ground). From the screen: amplitude, period (f = 1/T), phase difference, rise time, and waveform shape. Older CROs used an electron beam deflected by plates (thermionic electron emission from a heated cathode); digital storage scopes sample the signal and show it on an LCD.

Worked example: reading a trace

A sine wave covers 3.2 divisions peak to peak at 2 V/div, and one cycle spans 4.0 divisions at 0.5 ms/div.

  • Peak-to-peak = 6.4 V; peak = 3.2 V; r.m.s. = 3.2 ÷ √2 = 2.26 V.
  • Period = 2.0 ms; frequency = 500 Hz.

Other instruments

Signal (function) generators (sine, square, triangle at set frequency and amplitude); bench power supplies; frequency counters; LCR meters; logic probes and analysers for digital circuits.

Good practice and safety

Start on the highest range; never measure resistance in a live circuit; connect ammeters in series and voltmeters in parallel; use the probe's ×10 setting for high-frequency signals; never work on mains circuits without isolation and supervision.

Exams usually ask

  • Distinguish accuracy, precision and resolution, and types of error.
  • Calculate loading errors and design shunts and multipliers; interpret an oscilloscope trace.

Common mistake

Connecting an ammeter across a supply (in parallel): its very low resistance acts as a short circuit and blows its fuse.

Written by the UniHive team in our own words. Spotted a mistake? Tell us on WhatsApp.