Methods of hormone assay

Bioassay, radioimmunoassay, ELISA, chemiluminescence and mass spectrometry, standard curves and quality control.

Our notes

2 min read

Why special methods?

Hormones circulate at very low concentrations (pmol/L to nmol/L) among thousands of other molecules, so assays must be sensitive and specific.

Bioassays (historical)

Measure a biological effect (e.g. a pregnancy test using animal ovaries, or insulin's effect on blood glucose in rabbits). Slow, variable and ethically costly; now replaced.

Immunoassays

Use antibodies that bind the hormone specifically.

  • Radioimmunoassay (RIA) (Yalow and Berson; Nobel Prize 1977): a fixed amount of radioactively labelled hormone competes with the patient's hormone for a limited amount of antibody. More hormone in the sample → less labelled hormone bound → lower count. Very sensitive, but needs radioactivity.
  • ELISA (enzyme-linked immunosorbent assay): the label is an enzyme that makes a coloured or light-emitting product. Sandwich ELISA: a capture antibody on the plate binds the hormone; a second, enzyme-labelled antibody binds another part; colour is proportional to hormone concentration.
  • Chemiluminescence immunoassays: the label emits light; used in automated analysers in hospital labs.

Competitive and sandwich assays

TypeSignal vs concentrationUsed for
Competitivesignal falls as concentration risessmall molecules (steroids, T₄)
Sandwich (immunometric)signal rises with concentrationlarge proteins (TSH, LH, hCG, insulin)

Standard curves

Run standards of known concentration, plot signal against concentration, and read unknowns from the curve.

Worked example: reading a sandwich ELISA

Standards: 0 ng/mL → absorbance 0.05; 10 ng/mL → 0.45; 20 ng/mL → 0.85 (straight line in this range).

  • Slope = (0.85 − 0.05)/20 = 0.040 per ng/mL.
  • A sample reads 0.61: concentration = (0.61 − 0.05)/0.040 = 14 ng/mL.
  • If the sample was diluted 1 in 5 before the test, the original is 14 × 5 = 70 ng/mL.

Check: the two standards 10 and 20 give the same slope ((0.85 − 0.45)/10 = 0.040) ✓.

Mass spectrometry

Liquid chromatography with tandem mass spectrometry (LC-MS/MS) separates and identifies molecules by mass. It is more specific than immunoassays for steroids and vitamin D, and is used in reference and sports-doping laboratories.

Point-of-care tests

Lateral-flow strips (like home pregnancy tests detecting hCG in urine) use a sandwich immunoassay on paper with coloured particles.

Quality and pitfalls

  • Cross-reactivity: the antibody binds similar molecules.
  • Hook effect: extremely high levels can give falsely low results in some sandwich assays.
  • Interference: heterophile antibodies, biotin supplements in some kits.
  • Timing: cortisol and GH vary through the day.
  • Laboratories run controls of known value each batch and join external quality schemes.

Exams usually ask

  • Describe the principle of RIA or ELISA.
  • Compare competitive and sandwich assays.
  • Use a standard curve to find a concentration.

Common mistake

Reading a competitive assay as if a higher signal meant more hormone. In competitive assays a lower signal means a higher concentration.

Written by the UniHive team in our own words. Spotted a mistake? Tell us on WhatsApp.