UNIHIVE

Study Hub UNIPORT FSB 203

FSB 203 · Biological Techniques

The light microscope: structure, use and care

Parts of the compound microscope, magnification and resolution, measuring cells with a graticule, and caring for the instrument.

Our notes

2 min read

Parts and what they do

PartJob
Eyepiece (ocular), usually 10×magnifies the image formed by the objective
Objectives (4×, 10×, 40×, 100× oil) on a revolving nosepieceform the first magnified image
Stage with clips or mechanical stageholds the slide
Coarse and fine focus knobsmove the stage or tube to focus
Condenser and iris diaphragmfocus and control the light on the specimen
Light source or mirrorlights the specimen
Arm and basecarrying and support

Magnification and resolution

  • Total magnification = eyepiece × objective: 10 × 40 = 400×.
  • Resolution is the smallest distance between two points that can still be seen as separate: d = 0.61λ/NA (NA = numerical aperture of the objective).

Worked example. λ = 550 nm, NA = 1.25 (oil immersion): d = 0.61 × 550 ÷ 1.25 = 268 nm ≈ 0.27 μm. That is the limit of light microscopy: bacteria are visible, viruses are not. Magnifying further only gives "empty magnification".

  • Oil immersion: a drop of immersion oil (refractive index close to glass) between slide and the 100× lens collects more light and raises NA.

Using it

  1. Carry with one hand on the arm and the other under the base.
  2. Start on the lowest power; centre and focus with the coarse knob, watching from the side as you bring the lens close, then focus upward.
  3. Turn to higher power; use only the fine focus.
  4. Adjust the diaphragm: more light for high power, less for pale unstained specimens (more contrast).
  5. Draw what you see, with a scale bar and the magnification.

Measuring cells

An eyepiece graticule has a scale with no fixed value; calibrate it against a stage micrometer (usually 1 mm in 100 divisions, so 10 μm each).
Worked example. At 400×, 10 stage-micrometer divisions (100 μm) line up with 25 eyepiece units. One eyepiece unit = 100 ÷ 25 = 4 μm. A cell 12 eyepiece units long is 12 × 4 = 48 μm. Recalibrate for every objective.
Field of view: if the field is 1.8 mm across at 100×, at 400× it is 1.8 × 100/400 = 0.45 mm.

Care

  • Clean lenses only with lens tissue (never paper towels or your shirt); wipe oil off the 100× lens after use.
  • Never let a high-power objective touch a slide or liquid.
  • Lower the stage, set the lowest objective, switch off and cover with the dust cover. In humid Port Harcourt weather, fungus can grow on lenses: store with silica gel where possible.

Other microscopes

Phase contrast (live unstained cells), dark field (spirochaetes), fluorescence, and electron microscopes (TEM for thin sections, SEM for surfaces) with resolution near 1 nm.

Exams usually ask: label a microscope and give each part's function; calculate magnification, size from a graticule and field of view; explain resolution and oil immersion; list rules for care.

Common mistake: focusing downward on high power while looking through the eyepiece. That is how slides crack and lenses scratch.

Written by the UniHive team in our own words. Spotted a mistake? Tell us on WhatsApp.