FSB 203 · Biological Techniques

Colorimetry and spectrophotometry

Absorbance and transmittance, the Beer-Lambert law, standard curves, choosing a wavelength and using a spectrophotometer.

Our notes

2 min read

Principle

A coloured solution absorbs some wavelengths of light. The more of the substance there is, the more light it absorbs. A colorimeter uses coloured filters to choose the light; a spectrophotometer uses a prism or grating to choose a precise wavelength, including UV.

Transmittance and absorbance

  • Transmittance T = I/I₀ (light out ÷ light in), often as a percentage.
  • Absorbance A = log₁₀(I₀/I) = −log T.

T = 100% gives A = 0; T = 10% gives A = 1; T = 1% gives A = 2.

Beer-Lambert law

A = εcl

  • ε: molar absorptivity (L mol⁻¹ cm⁻¹), fixed for a substance at one wavelength.
  • c: concentration (mol L⁻¹).
  • l: path length (cm), usually 1 cm in a standard cuvette.

Worked example. NADH absorbs at 340 nm with ε = 6,220 L mol⁻¹ cm⁻¹. A sample in a 1 cm cuvette reads A = 0.311. c = A ÷ (εl) = 0.311 ÷ 6,220 = 5.0 × 10⁻⁵ mol L⁻¹ (50 μM). Check: 6,220 × 5.0 × 10⁻⁵ = 0.311.
The law holds best for dilute solutions, roughly A between 0.1 and 1.0. Dilute samples that read too high and multiply back.

The standard (calibration) curve

When ε is not known, prepare standards of known concentration, treat them like the samples (for example with a colour reagent), and plot A against concentration. Read unknowns from the line.
Example. Glucose standards 0, 2, 4, 6 and 8 mmol L⁻¹ give absorbances 0.00, 0.12, 0.24, 0.36, 0.48: a straight line through zero with slope 0.06 per mmol L⁻¹. A blood sample reading 0.33 contains 0.33 ÷ 0.06 = 5.5 mmol L⁻¹. Most clinical chemistry tests (glucose, protein, cholesterol) work this way.

Choosing the wavelength

Use the wavelength of maximum absorbance (λmax), found by scanning: the reading is most sensitive there and changes little with small wavelength errors. A substance absorbs the colour complementary to the one it shows: a red solution absorbs green-blue light.

Using the instrument

  1. Warm up the lamp; set the wavelength.
  2. Zero (100% T) with a blank: the solvent and reagents without the substance.
  3. Handle cuvettes by the frosted sides; wipe the clear sides; no bubbles or fingerprints. Use quartz cuvettes for UV (below about 340 nm), because glass and plastic absorb UV.
  4. Read standards and samples; re-check the blank now and then.

Uses in biology

Measuring protein (Biuret, Bradford), DNA (A₂₆₀; the ratio A₂₆₀/A₂₈₀ near 1.8 suggests pure DNA), enzyme reaction rates, bacterial growth (turbidity at 600 nm), chlorophyll content, and water quality (nitrate, phosphate).

Exams usually ask: define absorbance and transmittance; state and use the Beer-Lambert law; explain blanks and standard curves; explain the choice of wavelength.

Common mistake: reading samples outside the range of the standards. Extrapolating a standard curve is unreliable: dilute the sample.