FSB 203 · Biological Techniques
Our notes
2 min readPrinciple
A coloured solution absorbs some wavelengths of light. The more of the substance there is, the more light it absorbs. A colorimeter uses coloured filters to choose the light; a spectrophotometer uses a prism or grating to choose a precise wavelength, including UV.
Transmittance and absorbance
- Transmittance T = I/I₀ (light out ÷ light in), often as a percentage.
- Absorbance A = log₁₀(I₀/I) = −log T.
T = 100% gives A = 0; T = 10% gives A = 1; T = 1% gives A = 2.
Beer-Lambert law
A = εcl
- ε: molar absorptivity (L mol⁻¹ cm⁻¹), fixed for a substance at one wavelength.
- c: concentration (mol L⁻¹).
- l: path length (cm), usually 1 cm in a standard cuvette.
Worked example. NADH absorbs at 340 nm with ε = 6,220 L mol⁻¹ cm⁻¹. A sample in a 1 cm cuvette reads A = 0.311. c = A ÷ (εl) = 0.311 ÷ 6,220 = 5.0 × 10⁻⁵ mol L⁻¹ (50 μM). Check: 6,220 × 5.0 × 10⁻⁵ = 0.311.
The law holds best for dilute solutions, roughly A between 0.1 and 1.0. Dilute samples that read too high and multiply back.
The standard (calibration) curve
When ε is not known, prepare standards of known concentration, treat them like the samples (for example with a colour reagent), and plot A against concentration. Read unknowns from the line.
Example. Glucose standards 0, 2, 4, 6 and 8 mmol L⁻¹ give absorbances 0.00, 0.12, 0.24, 0.36, 0.48: a straight line through zero with slope 0.06 per mmol L⁻¹. A blood sample reading 0.33 contains 0.33 ÷ 0.06 = 5.5 mmol L⁻¹. Most clinical chemistry tests (glucose, protein, cholesterol) work this way.
Choosing the wavelength
Use the wavelength of maximum absorbance (λmax), found by scanning: the reading is most sensitive there and changes little with small wavelength errors. A substance absorbs the colour complementary to the one it shows: a red solution absorbs green-blue light.
Using the instrument
- Warm up the lamp; set the wavelength.
- Zero (100% T) with a blank: the solvent and reagents without the substance.
- Handle cuvettes by the frosted sides; wipe the clear sides; no bubbles or fingerprints. Use quartz cuvettes for UV (below about 340 nm), because glass and plastic absorb UV.
- Read standards and samples; re-check the blank now and then.
Uses in biology
Measuring protein (Biuret, Bradford), DNA (A₂₆₀; the ratio A₂₆₀/A₂₈₀ near 1.8 suggests pure DNA), enzyme reaction rates, bacterial growth (turbidity at 600 nm), chlorophyll content, and water quality (nitrate, phosphate).
Exams usually ask: define absorbance and transmittance; state and use the Beer-Lambert law; explain blanks and standard curves; explain the choice of wavelength.
Common mistake: reading samples outside the range of the standards. Extrapolating a standard curve is unreliable: dilute the sample.
Written by the UniHive team in our own words. Spotted a mistake? Tell us on WhatsApp.
Read
-
Best pick Beer–Lambert law
(opens en.wikipedia.org in a new tab)
Wikipedia, Beer–Lambert law, CC BY-SA 4.0 · licence
-
NOUN BIO 204: Biological Techniques
NOUN course book (PDF, about 2.1 MB). Microscope, biological drawing, dissection, slide preparation, microtomy, colorimetry, photometry, chromatography.
(opens nou.edu.ng in a new tab)
-
Spectrophotometry
(opens en.wikipedia.org in a new tab)
Wikipedia, Spectrophotometry, CC BY-SA 4.0 · licence
-
Colorimeter (chemistry)
(opens en.wikipedia.org in a new tab)
Wikipedia, Colorimeter (chemistry), CC BY-SA 4.0 · licence
Watch
-
Spectrophotometry and Beer's Law
6 min · ≈ 27 MB · about ₦15
(opens youtube.com in a new tab)
-
Spectrophotometry and the Beer–Lambert Law | AP Chemistry | Khan Academy
10 min · ≈ 45 MB · about ₦25
(opens youtube.com in a new tab)
-
Beer Lambert's Law, Absorbance & Transmittance - Spectrophotometry, Basic Introduction - Chemistry
18 min · ≈ 81 MB · about ₦45
(opens youtube.com in a new tab)
We link to free material and never copy it. Every link was opened by our team before it went up; video data is a rough guide at 360p.