ENG 213.2 · Basic Electrical Engineering

Electrical measurements and measuring instruments

Errors, moving-coil and moving-iron meters, shunts and multipliers, bridges, multimeters, wattmeters and oscilloscopes.

Our notes

3 min read

Measurement words

  • Accuracy: closeness to the true value. Precision: closeness of repeated readings.
  • Sensitivity: output change per input change.
  • Resolution: smallest change that can be detected.
  • Errors: gross (human: misreading, parallax), systematic (instrument, calibration, environment, loading) and random (noise; reduced by averaging).
  • Limiting error is often given as a % of full scale. A 100 V meter accurate to ±1% of full scale reading 25 V has an error of ±1 V, which is ±4% of the reading: read meters near the top of the scale.

Indicating instruments

Every analogue instrument needs a deflecting torque, a controlling torque (springs) and damping (to stop the pointer oscillating).

  • Permanent-magnet moving-coil (PMMC): a coil in a magnet's field; deflection proportional to current (uniform scale); accurate and sensitive; DC only (with a rectifier for AC).
  • Moving-iron: iron vanes in a coil's field; works on AC and DC; non-uniform scale; robust and cheap (switchboards).
  • Electrodynamometer: fixed and moving coils; used as the wattmeter.
  • Induction type: the old rotating-disc energy meter (now mostly replaced by electronic and prepaid meters).

Extending a meter's range

Example meter: full-scale current I_m = 10 mA, resistance R_m = 5 Ω (full-scale voltage 0.05 V).

  • Ammeter to 1 A: add a shunt in parallel: R_sh = I_m R_m/(I − I_m) = 0.05/0.99 ≈ 0.0505 Ω. Check: at 1 A, 0.99 A through the shunt gives 0.99 × 0.0505 ≈ 0.05 V ✓.
  • Voltmeter to 100 V: add a multiplier in series: R = V/I_m − R_m = 100/0.01 − 5 = 9,995 Ω. Its sensitivity is 1/I_m = 100 Ω/V.

Loading effect

A voltmeter draws current and changes the circuit. Measuring across 10 kΩ in a divider (two 10 kΩ on 20 V) with a 10 kΩ meter: the parallel combination is 5 kΩ and the reading is 20 × 5/15 ≈ 6.7 V instead of 10 V. Use meters with resistance much higher than the circuit (digital meters: about 10 MΩ). Ammeters must have very low resistance.

Measuring resistance

  • Ammeter-voltmeter method.
  • Wheatstone bridge: four arms; at balance (galvanometer reads zero) R_x = R₃ R₂/R₁ (R₁/R₂ = R₃/R_x). Accurate for medium resistances; also used in strain gauges.
  • Insulation tester ("megger"): applies 500 V or 1 kV to measure the insulation resistance (megohms) of cables and motors.
  • Earth tester: measures earth electrode resistance (safety of installations).

Electronic instruments

  • Digital multimeter (DMM): voltage, current, resistance, continuity, often frequency and capacitance; shows true RMS on better models.
  • Clamp meter: measures current without breaking the circuit (magnetic field of the wire).
  • Oscilloscope: shows a waveform against time: read peak voltage, period (frequency), phase difference and noise.

Example: a sine wave 4 divisions peak-to-peak at 5 V/div and 4 divisions per cycle at 5 ms/div: V_pp = 20 V (V_m = 10 V, V_rms ≈ 7.07 V), T = 20 ms, f = 50 Hz.

Exams usually ask

  • Design a shunt or multiplier for a moving-coil meter.
  • Compare PMMC and moving-iron instruments.
  • Derive the balance condition of the Wheatstone bridge; explain loading errors.

Common mistake

Connecting an ammeter in parallel across a load. Ammeters go in series; with their tiny resistance a parallel connection is a short circuit that can blow the meter's fuse.

Written by the UniHive team in our own words. Spotted a mistake? Tell us on WhatsApp.