CHM 362.2 · Environmental Chemistry
Our notes
2 min readSampling comes first
A perfect instrument cannot rescue a bad sample.
- Representative: the sample must stand for the whole (several points, depths and times).
- Containers: glass for oils and organics; plastic for most metals and ions.
- Preservation: cool to about 4°C; acidify metal samples with nitric acid; analyse BOD within a short time.
- Records: place, date, time, sampler, chain of custody.
Field measurements
| Parameter | Instrument |
|---|---|
| pH | pH meter (glass electrode); calibrate with buffers |
| Conductivity | Conductivity meter: shows total dissolved ions |
| DO | DO meter or Winkler titration |
| Turbidity | Turbidimeter (NTU) |
| Temperature | Thermometer |
UV-visible spectrophotometry
Measures light absorbed by a coloured solution. Beer-Lambert law: A = εbc.
- A: absorbance; ε: molar absorptivity; b: path length; c: concentration.
Many pollutants (nitrate, phosphate, ammonia, iron) are reacted to form a coloured complex first.
Worked example
A phosphate complex has ε = 15,000 L mol⁻¹ cm⁻¹. In a 1 cm cell, A = 0.45.
c = A ÷ (εb) = 0.45 ÷ 15,000 = 3.0 × 10⁻⁵ mol/L.
Check: 15,000 × 1 × 3.0 × 10⁻⁵ = 0.45. Correct.
Calibration curves
In practice, measure standards of known concentration, plot A against c, and read the sample from the line. Check the blank and stay within the straight-line range.
Atomic absorption spectroscopy (AAS)
- The sample is atomised in a flame (or graphite furnace).
- A lamp of the element being measured shines through; atoms absorb their own wavelength.
- Used for lead, cadmium, zinc, copper, iron, chromium.
- One element at a time.
ICP-OES and ICP-MS
A very hot argon plasma excites or ionises atoms. ICP-OES measures emitted light; ICP-MS measures ions by mass. Many elements at once; ICP-MS reaches very low levels.
Chromatography
| Technique | For |
|---|---|
| Gas chromatography (GC) | Volatile organics: petroleum hydrocarbons, pesticides, BTEX |
| GC-MS | Identifies compounds by their mass spectra |
| HPLC | Non-volatile organics, PAHs |
| Ion chromatography | Anions: nitrate, sulphate, chloride |
Air monitoring
- Particles: high-volume samplers with filters weighed before and after; low-cost optical sensors (less accurate).
- Gases: electrochemical cells, chemiluminescence (NOx), UV fluorescence (SO₂), UV absorption (O₃).
Quality assurance
- Blanks, duplicates and certified reference materials.
- Detection limit: the lowest level reliably detected.
- Accredited laboratories and methods (such as standard methods for water and wastewater).
Exams usually ask
- State the Beer-Lambert law and calculate a concentration.
- Name the instrument suited to metals, volatile organics and anions.
Common mistake
Storing a water sample for metals in an unwashed bottle with no acid. Metals stick to the walls or come from the container.
Written by the UniHive team in our own words. Spotted a mistake? Tell us on WhatsApp.
Read
-
Atomic absorption spectroscopy
(opens en.wikipedia.org in a new tab)
Wikipedia, Atomic absorption spectroscopy, CC BY-SA 4.0 · licence
-
Gas chromatography
(opens en.wikipedia.org in a new tab)
Wikipedia, Gas chromatography, CC BY-SA 4.0 · licence
-
Ultraviolet–visible spectroscopy
(opens en.wikipedia.org in a new tab)
Wikipedia, Ultraviolet–visible spectroscopy, CC BY-SA 4.0 · licence
-
Environmental monitoring
(opens en.wikipedia.org in a new tab)
Wikipedia, Environmental monitoring, CC BY-SA 4.0 · licence
Watch
-
Best pick Atomic Absorption Spectroscopy (AAS)
11 min · ≈ 50 MB · about ₦28
(opens youtube.com in a new tab)
-
Atomic Absorption Spectroscopy Part 1
15 min · ≈ 68 MB · about ₦38
(opens youtube.com in a new tab)
-
Atomic Absorption Spectroscopy || Ch 3 Unit 2 || Instrumental method of Analysis 7th Semester || CWP
25 min · ≈ 113 MB · about ₦63
(opens youtube.com in a new tab)
-
Atomic Absorption Instrument (Spectroscopy)
47 min · ≈ 212 MB · about ₦119
(opens youtube.com in a new tab)
We link to free material and never copy it. Every link was opened by our team before it went up; video data is a rough guide at 360p.