Sampling and instrumental analysis

Taking good samples, and the instruments used to measure pollutants: UV-visible, AAS, ICP, chromatography and field meters.

Our notes

2 min read

Sampling comes first

A perfect instrument cannot rescue a bad sample.

  • Representative: the sample must stand for the whole (several points, depths and times).
  • Containers: glass for oils and organics; plastic for most metals and ions.
  • Preservation: cool to about 4°C; acidify metal samples with nitric acid; analyse BOD within a short time.
  • Records: place, date, time, sampler, chain of custody.

Field measurements

ParameterInstrument
pHpH meter (glass electrode); calibrate with buffers
ConductivityConductivity meter: shows total dissolved ions
DODO meter or Winkler titration
TurbidityTurbidimeter (NTU)
TemperatureThermometer

UV-visible spectrophotometry

Measures light absorbed by a coloured solution. Beer-Lambert law: A = εbc.

  • A: absorbance; ε: molar absorptivity; b: path length; c: concentration.

Many pollutants (nitrate, phosphate, ammonia, iron) are reacted to form a coloured complex first.

Worked example

A phosphate complex has ε = 15,000 L mol⁻¹ cm⁻¹. In a 1 cm cell, A = 0.45.
c = A ÷ (εb) = 0.45 ÷ 15,000 = 3.0 × 10⁻⁵ mol/L.
Check: 15,000 × 1 × 3.0 × 10⁻⁵ = 0.45. Correct.

Calibration curves

In practice, measure standards of known concentration, plot A against c, and read the sample from the line. Check the blank and stay within the straight-line range.

Atomic absorption spectroscopy (AAS)

  • The sample is atomised in a flame (or graphite furnace).
  • A lamp of the element being measured shines through; atoms absorb their own wavelength.
  • Used for lead, cadmium, zinc, copper, iron, chromium.
  • One element at a time.

ICP-OES and ICP-MS

A very hot argon plasma excites or ionises atoms. ICP-OES measures emitted light; ICP-MS measures ions by mass. Many elements at once; ICP-MS reaches very low levels.

Chromatography

TechniqueFor
Gas chromatography (GC)Volatile organics: petroleum hydrocarbons, pesticides, BTEX
GC-MSIdentifies compounds by their mass spectra
HPLCNon-volatile organics, PAHs
Ion chromatographyAnions: nitrate, sulphate, chloride

Air monitoring

  • Particles: high-volume samplers with filters weighed before and after; low-cost optical sensors (less accurate).
  • Gases: electrochemical cells, chemiluminescence (NOx), UV fluorescence (SO₂), UV absorption (O₃).

Quality assurance

  • Blanks, duplicates and certified reference materials.
  • Detection limit: the lowest level reliably detected.
  • Accredited laboratories and methods (such as standard methods for water and wastewater).

Exams usually ask

  • State the Beer-Lambert law and calculate a concentration.
  • Name the instrument suited to metals, volatile organics and anions.

Common mistake

Storing a water sample for metals in an unwashed bottle with no acid. Metals stick to the walls or come from the container.

Written by the UniHive team in our own words. Spotted a mistake? Tell us on WhatsApp.