Electroanalytical methods

These methods use electrical measurements (current, charge, potential) to analyse. Two of the oldest are electrogravimetry and coulometry, both based on electrolysis: driving a non-spontaneous reaction with an external power supply.

Electrogravimetry

The analyte metal is deposited on a weighed electrode (usually a platinum gauze cathode) by electrolysis, then the electrode is washed, dried and weighed again. The gain in mass is the metal.
Example: copper in brass: Cu²⁺ + 2e⁻ → Cu(s) at the cathode; at the anode water is oxidised to O₂.
Requirements:

  • The deposit must be adherent, dense and smooth so it does not flake off when washed. Helped by moderate current, stirring and suitable additives (nitric acid in copper deposition helps prevent hydrogen bubbles).
  • Deposition must be complete (test: no further mass gain, or no metal in a test drop).

Two modes:

  • Constant current: fast but not selective: as Cu²⁺ runs low, the potential rises and other metals or hydrogen may deposit.
  • Controlled (constant) potential: the cathode potential is held so that only the target metal deposits: selective (copper can be separated from tin and lead).

Faraday's laws

  1. The mass of substance changed at an electrode is proportional to the charge passed.
  2. For the same charge, masses are proportional to molar mass divided by the number of electrons per ion.

Charge Q = I × t (coulombs = amperes × seconds). Moles of electrons = Q / F, where F = 96,485 C/mol. Moles of substance = Q / (nF), where n is electrons per ion.

Coulometry

Measures the charge needed to convert the analyte completely. No weighing and no standards are needed: the electron is the "reagent", and charge can be measured very accurately. 100% current efficiency is essential: all the current must go to the analyte reaction.

  • Controlled-potential (potentiostatic) coulometry: potential held constant; current falls toward zero as analyte is used up; charge found by integrating current over time.
  • Constant-current coulometry (coulometric titration): a titrant is generated electrically at constant current (for example Br₂ from Br⁻, or I₂ from I⁻, or OH⁻ at a cathode); the time to the endpoint gives the charge. Good for small amounts and for unstable titrants. The Karl Fischer coulometric method for traces of water in oils and drugs works this way.

Worked example

A copper solution is electrolysed at a constant 0.500 A for 1,930 s, with 100% current efficiency.
Q = 0.500 × 1,930 = 965 C. Moles of electrons = 965 / 96,485 = 0.01000 mol. Cu²⁺ needs 2 electrons, so moles Cu = 0.00500 mol. Mass Cu = 0.00500 × 63.55 = 0.318 g.
In electrogravimetry this would be checked by weighing: the cathode gains 0.318 g.

Exams usually ask

  • Describe electrogravimetry and the conditions for a good deposit.
  • State Faraday's laws and use Q = It.
  • Distinguish controlled-potential and constant-current coulometry.

Common mistake

Forgetting n, the number of electrons per ion. Copper(II) needs two electrons per atom, so the moles of copper are half the moles of electrons.

Written by the UniHive team in our own words. Spotted a mistake? Tell us on WhatsApp.