Calibration curves, correlation and regression

Standards and calibration, the least-squares line, the correlation coefficient and finding an unknown.

Our notes

2 min read

Calibration

Most instrumental methods give a signal (absorbance, current, peak area) that must be converted to concentration. A calibration curve is made by measuring the signal for several standards of known concentration, plotting signal (y) against concentration (x), and fitting a line. An unknown's concentration is then read from its signal.
Good practice: at least five standards spread over the working range, a blank (zero standard), standards in a matrix like the samples, and unknowns inside the range (do not extrapolate).

The least-squares line

For a straight line y = mx + c, the best-fit slope and intercept minimise the sum of squared vertical deviations:

  • m = Σ(xᵢ − x̄)(yᵢ − ȳ) / Σ(xᵢ − x̄)²
  • c = ȳ − m x̄

The correlation coefficient

r = Σ(xᵢ − x̄)(yᵢ − ȳ) / √[Σ(xᵢ − x̄)² × Σ(yᵢ − ȳ)²]
r ranges from −1 to +1. Values close to ±1 show a strong linear relationship; calibration lines usually have r above 0.999. r measures linearity, not accuracy, and a high r can hide curvature: always look at the plot (and the residuals).

Worked example

Iron standards measured by spectrophotometry:

Concentration x (ppm)02468
Absorbance y0.0020.1620.3180.4810.640
  1. x̄ = 4; ȳ = 1.603 / 5 = 0.3206.
  2. Σ(x − x̄)² = 16 + 4 + 0 + 4 + 16 = 40.
  3. Σ(x − x̄)(y − ȳ) = (−4)(−0.3186) + (−2)(−0.1586) + 0 + 2(0.1604) + 4(0.3194) = 1.2744 + 0.3172 + 0.3208 + 1.2776 = 3.190.
  4. m = 3.190 / 40 = 0.07975 per ppm; c = 0.3206 − 0.07975 × 4 = 0.0016.
  5. r = 0.99998 (very linear).
  6. An unknown gives A = 0.400: x = (0.400 − 0.0016) / 0.07975 = 5.00 ppm.

(Check: 0.07975 × 5.00 + 0.0016 = 0.4004 ✓.)

Beer-Lambert law (why the line is straight)

In absorption spectrophotometry, A = εbc (ε molar absorptivity, b path length, c concentration), so absorbance is proportional to concentration over a working range. At high concentration the line curves.

Standard addition and internal standards

  • Standard addition: known amounts of analyte are added to portions of the sample itself; corrects for matrix effects.
  • Internal standard: a fixed amount of another substance added to all samples and standards; corrects for variations in the instrument or injection.

Exams usually ask

  • Calculate the slope, intercept and r for a calibration data set.
  • Use the line to find an unknown concentration.
  • Explain why extrapolation beyond the standards is unsafe.

Common mistake

Forcing the line through the origin when the blank is not zero, or reading an unknown far outside the standards' range.

Written by the UniHive team in our own words. Spotted a mistake? Tell us on WhatsApp.