BCH 511 · Industrial Biochemistry
Our notes
2 min readWhy diagnostic kits
Clinics need tests that are fast, cheap, accurate and stable without a full laboratory: blood glucose, malaria, HIV, hepatitis B, pregnancy, typhoid, urinalysis. Most kits used in Nigeria are imported; local production saves foreign exchange and suits local needs (heat-stable formats).
Enzyme-coupled colorimetric assays
An enzyme converts the analyte; a second reaction makes a colour measured by a photometer (Beer-Lambert law: absorbance ∝ concentration).
Glucose (GOD-POD, Trinder method):
glucose + O₂ → (glucose oxidase) gluconic acid + H₂O₂;
H₂O₂ + phenol + 4-aminoantipyrine → (peroxidase) pink quinoneimine dye, read at about 505 nm.
Worked example. Standard (100 mg/dL) gives absorbance 0.40; a patient's serum gives 0.58. Glucose = 0.58/0.40 × 100 = 145 mg/dL (8.1 mmol/L: divide mg/dL by 18).
Similar kits: cholesterol (cholesterol oxidase), triglycerides, urea (urease), creatinine, uric acid (uricase), ALT/AST (liver enzymes), bilirubin.
Components: buffer, enzymes, chromogen, stabilisers, preservative, a standard (calibrator) and controls.
Dipsticks and test strips
Reagents dried on paper pads: urine strips for glucose (GOD-POD), protein (tetrabromophenol blue protein error of indicators), pH, blood (peroxidase-like activity of haem), ketones (nitroprusside), nitrite and leukocytes. Read by colour chart.
Immunoassays
Use antibodies' specific binding.
- ELISA: antigen or antibody bound to a plate; enzyme-labelled antibody gives colour with substrate (horseradish peroxidase + TMB). Formats: direct, indirect (detects antibodies, as in HIV screening), sandwich (detects antigens, as in hepatitis B surface antigen).
- Lateral-flow rapid tests (RDTs): sample flows along a strip, binds gold- or latex-labelled antibodies, and a coloured line forms at the test line; a control line shows the test worked. Malaria RDTs detect HRP2 (P. falciparum) or pLDH antigens; pregnancy tests detect hCG.
- Agglutination tests: latex or red cells clump in the presence of antigen/antibody (blood grouping, Widal test for typhoid: widely used in Nigeria but has poor specificity; culture or newer tests are better).
Biosensors
Biological element + transducer. Glucometers use glucose oxidase or glucose dehydrogenase on an electrode strip; the current is proportional to glucose (amperometric). Continuous glucose monitors use the same idea under the skin.
Making a kit: quality
- Sensitivity = true positives / all who have the disease; specificity = true negatives / all who do not.
Worked example. A malaria RDT tested on 200 people (100 confirmed infected by microscopy): 92 positive among the infected, 6 positive among the 100 uninfected. Sensitivity = 92/100 = 92%; specificity = 94/100 = 94%.
- Also: precision (repeatability), accuracy against a reference method, limit of detection, stability (accelerated ageing at raised temperatures to predict shelf life), lot-to-lot consistency.
- Regulation: NAFDAC registers in vitro diagnostics; medical laboratory practice is regulated by the Medical Laboratory Science Council of Nigeria; WHO prequalification guides procurement of HIV and malaria tests.
Exams usually ask: principle of the GOD-POD glucose assay and calculating results; immunoassay formats (ELISA, lateral flow, agglutination); how a glucometer works; sensitivity and specificity calculations; requirements for producing and validating a diagnostic kit.
Common mistake: mixing up sensitivity and specificity. Sensitivity is about catching the sick (few false negatives); specificity is about clearing the well (few false positives).
Written by the UniHive team in our own words. Spotted a mistake? Tell us on WhatsApp.
Read
-
Best pick Microbiology, 20.4 EIAs and ELISAs
(opens openstax.org in a new tab)
OpenStax, Microbiology, 20.4 EIAs and ELISAs, CC BY-NC-SA 4.0 · licence
-
Microbiology, 20.1 Polyclonal and Monoclonal Antibody Production
(opens openstax.org in a new tab)
OpenStax, Microbiology, 20.1 Polyclonal and Monoclonal Antibody Production, CC BY-NC-SA 4.0 · licence
-
Lateral flow test
(opens en.wikipedia.org in a new tab)
Wikipedia, Lateral flow test, CC BY-SA 4.0 · licence
-
Glucose meter
(opens en.wikipedia.org in a new tab)
Wikipedia, Glucose meter, CC BY-SA 4.0 · licence
-
Malaria antigen detection tests
(opens en.wikipedia.org in a new tab)
Wikipedia, Malaria antigen detection tests, CC BY-SA 4.0 · licence
Watch
We link to free material and never copy it. Every link was opened by our team before it went up; video data is a rough guide at 360p.