2 min readWhy special methods?
Hormones circulate at very low concentrations (pmol/L to nmol/L) among thousands of other molecules, so assays must be sensitive and specific.
Bioassays (historical)
Measure a biological effect (e.g. a pregnancy test using animal ovaries, or insulin's effect on blood glucose in rabbits). Slow, variable and ethically costly; now replaced.
Immunoassays
Use antibodies that bind the hormone specifically.
- Radioimmunoassay (RIA) (Yalow and Berson; Nobel Prize 1977): a fixed amount of radioactively labelled hormone competes with the patient's hormone for a limited amount of antibody. More hormone in the sample → less labelled hormone bound → lower count. Very sensitive, but needs radioactivity.
- ELISA (enzyme-linked immunosorbent assay): the label is an enzyme that makes a coloured or light-emitting product. Sandwich ELISA: a capture antibody on the plate binds the hormone; a second, enzyme-labelled antibody binds another part; colour is proportional to hormone concentration.
- Chemiluminescence immunoassays: the label emits light; used in automated analysers in hospital labs.
Competitive and sandwich assays
| Type | Signal vs concentration | Used for |
|---|---|---|
| Competitive | signal falls as concentration rises | small molecules (steroids, T₄) |
| Sandwich (immunometric) | signal rises with concentration | large proteins (TSH, LH, hCG, insulin) |
Standard curves
Run standards of known concentration, plot signal against concentration, and read unknowns from the curve.
Worked example: reading a sandwich ELISA
Standards: 0 ng/mL → absorbance 0.05; 10 ng/mL → 0.45; 20 ng/mL → 0.85 (straight line in this range).
- Slope = (0.85 − 0.05)/20 = 0.040 per ng/mL.
- A sample reads 0.61: concentration = (0.61 − 0.05)/0.040 = 14 ng/mL.
- If the sample was diluted 1 in 5 before the test, the original is 14 × 5 = 70 ng/mL.
Check: the two standards 10 and 20 give the same slope ((0.85 − 0.45)/10 = 0.040) ✓.
Mass spectrometry
Liquid chromatography with tandem mass spectrometry (LC-MS/MS) separates and identifies molecules by mass. It is more specific than immunoassays for steroids and vitamin D, and is used in reference and sports-doping laboratories.
Point-of-care tests
Lateral-flow strips (like home pregnancy tests detecting hCG in urine) use a sandwich immunoassay on paper with coloured particles.
Quality and pitfalls
- Cross-reactivity: the antibody binds similar molecules.
- Hook effect: extremely high levels can give falsely low results in some sandwich assays.
- Interference: heterophile antibodies, biotin supplements in some kits.
- Timing: cortisol and GH vary through the day.
- Laboratories run controls of known value each batch and join external quality schemes.
Exams usually ask
- Describe the principle of RIA or ELISA.
- Compare competitive and sandwich assays.
- Use a standard curve to find a concentration.
Common mistake
Reading a competitive assay as if a higher signal meant more hormone. In competitive assays a lower signal means a higher concentration.
Written by the UniHive team in our own words. Spotted a mistake? Tell us on WhatsApp.